Connect the cables labeled I/O. On the P360+, connect the cables to the inputs labeled I/O Coil A, B & C (I, O). On the P260+, connect the cables to the inputs labeled I/O Coil (I, O).
P360+ electronic control output connections.P260+ electronic control output connections.ACTAS P360+ with control outputs for three closing and three opening coils. ACTAS P260+ with control outputs for one closing and two opening coils.
Electronic control outputs (IGBTs), with which DC and AC voltage coils can be operated, are used to operate the switching coils.
The coil operating currents are measured using shunts integrated in the test system, with a maximum measuring range of 100 A.
2
Follow the connection example
Connection example on the plus or ground branch of the operating coil.
Operating-coil connection example.
Connection procedure
Relay Output Connections
1
Connect the relay cables
Connect the cables labeled Relay. On the P360+, connect the cables to the inputs labeled Ralais (1, 2). On the P260+, connect the cables to the inputs labeled Ralais.
ACTAS P360+ with two relais outputs.ACTAS P260+ with one relais output.
The relay outputs are controlled via the software just like the IGBT control outputs. Switch-on and switch-off times can be set relative to the start and end time of the test; see Settings for the switching device/Relays.